uScopeDX

uScopeDX ModelsuScopeDX-2040
Product Information
uScopeDX-2060
Product Information
uScopeDX-4060
Product Information
Objective120X (0.40NA)
160mm Conjugate
40X (0.65NA)
160mm Conjugate
20X (0.40NA)
160mm Conjugate
60X (0.85NA)
160mm Conjugate
40X (0.65NA)
160mm Conjugate
60X (0.85NA)
160mm Conjugate
Camera2MP CMOS, High Speed2MP CMOS, High Speed2MP CMOS, High Speed
Camera Modes2
SD: 960×540
HD: 1920×1080
SD: 960×540
HD: 1920×1080
SD: 960×540
HD: 1920×1080
Pixel Mapping (SD)0.50µm/pixel0.25µm/pixel0.50µm/pixel0.167µm/pixel0.25µm/pixel0.167µm/pixel
Pixel Mapping (HD)0.25µm/pixel0.125µm/pixel0.25µm/pixel0.083µm/pixel0.125µm/pixel0.083µm/pixel
Image Resolution (SD)50,000 DPI100,000 DPI50,000 DPI150,000 DPI100,000 DPI150,000 DPI
Image Resolution (HD)100,000 DPI200,000 DPI100,000 DPI300,000 DPI200,000 DPI300,000 DPI
Field of View (mm)0.48×0.270.24×0.130.48×0.270.16×0.090.24×0.130.16×0.09
Slide Tray
Removable, Various Sizes
Dual 1" × 3" (included)
Single 2" × 3" (optional)
Removable, Various Sizes
Dual 1" × 3" (included)
Single 2" × 3" (optional)
Removable, Various Sizes
Dual 1" × 3" (included)
Single 2" × 3" (optional)
Slide CapacityDual or Single SlideDual or Single SlideDual or Single Slide
Scan Mode
Normal (AutoFocus)
Z-Stacking3
EDF4 (Ext. Depth-of-Field)
Normal (AutoFocus)
Z-Stacking3
EDF4 (Ext. Depth-of-Field)
Normal (AutoFocus)
Z-Stacking3
EDF4 (Ext. Depth-of-Field)
Focus Method5
Fixed Focus (Initial Only)
Fast Stack
Exhaustive Stack
Predictive
Predictive Stack
Fixed Focus (Initial Only)
Fast Stack
Exhaustive Stack
Predictive
Predictive Stack
Fixed Focus (Initial Only)
Fast Stack
Exhaustive Stack
Predictive
Predictive Stack
Scan Time6
SD: 5:30
HD: 8:00
SD: 22:00
HD: 31:00
SD: 5:30
HD: 8:00
SD: 49:30
HD: 1:10:00
SD: 22:00
HD: 31:00
SD: 49:30
HD: 1:10:00
Included SoftwareuScope Navigator v4uScope Navigator v4uScope Navigator v4
Optional SoftwareuScope EDF4 v1uScope EDF4 v1uScope EDF4 v1
Scan Field Image Format
Compressed (JPG Q90)
Uncompressed (PNG)
Compressed (JPG Q90)
Uncompressed (PNG)
Compressed (JPG Q90)
Uncompressed (PNG)
Whole Slide Image Format
Deep Zoom Pyramid
Tiled TIFF (SVS)
Easy Zoom (SZI)
Deep Zoom Pyramid
Tiled TIFF (SVS)
Easy Zoom (SZI)
Deep Zoom Pyramid
Tiled TIFF (SVS)
Easy Zoom (SZI)
Net Weight
15.5 lbs
7.0 kg
15.5 lbs
7.0 kg
15.5 lbs
7.0 kg
Dimensions7.75" × 6.57" × 10.63"
196mm × 167mm × 270mm
7.75" × 6.57" × 10.63"
196mm × 167mm × 270mm
7.75" × 6.57" × 10.63"
196mm × 167mm × 270mm
Brochure
 uScopeDX-2040
Product Information
uScopeDX-2060
Product Information
uScopeDX-4060
Product Information

1 The uScopeDX is a dual-objective whole slide scanner. Each uScopeDX model includes the specified objectives (20x, 40x, or 60x). Neither digital pixel doubling nor interpolation are used to achieve the specified magnification.

2 The camera can operate in one of two modes: HD where each pixel on the image sensor maps to one pixel on the captured image; or SD where a 2×2 array of pixels is used to generate one pixel on the captured image (this is often referred to as binning). HD mode provides higher resolution images (comparable to the next higher magnification objective) where SD mode provides better contrast and less camera noise.

3 Z-Stacking simply means that the scanner captures several images at different focus levels for each field of view. The uScope Navigator software composites these into a whole slide image with a focus adjustment that allows you to view each layer of the Z-stack using a natural control similar to the focus knob of a traditional microscope. Note: Z-Stacking requires significantly more time to scan a slide as many times more images are captured and saved to disk for each scan field.

4 EDF (extended depth-of-field) is an image processing technique that takes the best focus parts of each image from a Z-Stack and creates a new image where everything is in focus. This technique can be very beneficial for extremely thick sections where the natural depth-of-field of the objective is too shallow to show each discrete element in focus. Note: Optional software is required to create EDF whole slide images. Compositing EDF scan fields is a very processor-intensive operation that is best suited to multi-core CPUs.

5 The uScopeDX offers a number of different focus methods to accommodate a wide variety of specimen types and preparation. Specimens that are extremely thin can be scanned very well using predictive focus, fast stack, or even fixed focus. Thicker or inconsistent sections are best scanned with the exhaustive focus or predictive stack focus methods.

6 Scan speed is measured by scanning a 15mm×15mm area using the Predictive focus method. Fixed focus scans are slightly faster. Focus stack methods are slower (but may yield better focused images).

Contact Microscopes International or speak with your local distributor.
Copyright © 2024 Microscopes International, LLC. All rights reserved.