Support

uScope Navigator (Device Settings)

Overexposed Images after Changing Camera Mode
uScope Navigator (Device Settings)

Article ID: NDS1067
Created: 11 Aug 2017
Reviewed: 12 Nov 2017

Information in this article applies to:

  • uScope Navigator V4.0 — uScope Navigator V4.2.1

Symptom

After changing objective camera modes back and forth between HD and SD, images captured appear to be overexposed or to have strange exposure settings.

Cause

This problem is caused when the camera settings are not completely reset when changing camera modes. This problem does not consistently occur for all users.

Resolution

This problem has been corrected in uScope Navigator v4.3.

As a work-around for earlier releases, we suggest that after changing the camera mode, if image exposure seems odd, exit the uScope Navigator software, cycle power on the uScope, and re-launch the uScope Navigator application.