When creating a whole scan image, the resulting image has horizontal and/or vertical blank areas (gaps) between fields as shown.
Typically, whole scan images are composited from overlapped and stitched image fields and do not have gaps between the scan fields.
Fields in a scanned region of interest can be...
When scan fields are not overlapped (Tiled or Underlapped), they cannot be stitched. In such cases, they are rendered with a gutter (or gap) between the fields in the created whole scan image (as shown).
To overlap the fields and eliminate the gutter, change the Overlap Method for slide scans to Overlap to Stitch or to Custom and enter the number of pixels between fields both vertically (Y Overlap) and horizontally (X Overlap).